Radiofrequency coil

Radiofrequency coils (RF coils) are the receivers, and sometimes also the transmitters, of radiofrequency (RF) signals in equipment used in magnetic resonance imaging (MRI).

The MR signal in MRI is produced by the process of resonance, which is the result of radiofrequency coils. They consist of two electromagnetic coils, the transmitter and receiver, which generate the field and receive the resulting signal. Atomic nuclei of interest in MRI studies have their own resonant frequencies, in the radiofrequency portion of the electromagnetic spectrum.[1]

Although the electromagnetic fields produced by the transmitting coil are in the RF range of tens of megahertz (often in the shortwave radio portion of the electromagnetic spectrum) at powers usually exceeding the highest powers used by amateur radio, there is very little RF interference produced by the MRI machine. The reason for this is that the MRI is a very poor radio transmitter, and is without an antenna. The RF frequency electromagnetic field produced in the "transmitting coil" is a magnetic near-field with very little associated changing electric field component (such as all conventional radio wave transmissions have). Thus, the high-powered electromagnetic field produced in the MRI transmitter coil does not produce much electromagnetic radiation at its RF frequency, and the RF power is confined to the coil space and not radiated as "radio waves." Thus, the transmitting coil is a good EM near-field generator at radio frequency, but a poor EM radiation transmitter at radio frequency.

The receiver coil picks up the RF electromagnetic radiation produced by nuclear relaxation inside the subject. The signal acquired by the coil is an induced emf, and is not the result of picking up radio waves. This is a common misconception, and unfortunately, has propagated through the literature. MRI scanners are generally situated in metal mesh lined rooms which act as Faraday cages.)

Types of Radiofrequency Coil

RF coils for MRI can be grouped into two different classes: volume coils and surface coils.

Volume Coils

Volume coils are designed to provide a homogeneous RF excitation across a large volume. Most clinical MRI scanners include a built in volume coil to perform whole-body imaging, and smaller volume coils have been constructed for the head and other extremities.

Common designs for volume coils include Birdcage Coils, TEM Coils,[2] and Saddle Coils. These coils require a great deal of RF power because of their size, so they are often driven in quadrature in order to reduce by two the RF power requirements.

The RF homogeneity of volume coils is highly desirable for transmission, but is less ideal when the region of interest is small. The large field of view of volume coils means that they receive noise from the whole body, not just the region of interest.

Surface Coils

Surface coils are designed to provide a very high RF sensitivity over a small region of interest. These coils are often single or multi-turn loops which are placed directly over the anatomy of interest. The size of these coils can be optimized for the specific region of interest.

Surface coils make poor transmission coils because they have poor RF homogeneity, even over their region of interest. Their small field of view makes them ideal as receivers, as they only detect noise from the region of interest.

See also

References

  1. Huettel, S.A. Functional Magnetic Resonance Imaging. USA: Sinauer. p. 31.
  2. Vaughan, J.T.; Adriany, G.; Snyder, C.J.; Tian, J.; et al. (1 October 2004). "Efficient high-frequency body coil for high-field MRI". Magnetic Resonance in Medicine. 52 (4): 851–859. doi:10.1002/mrm.20177.
This article is issued from Wikipedia - version of the 10/11/2016. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.