Bioresorbable stent

In medicine, a stent is any device which is inserted into a blood vessel or other internal duct in order to expand the vessel to prevent or alleviate a blockage. Traditionally, such devices are fabricated from metal mesh and remain in the body permanently or until removed through further surgical intervention. A bioresorbable stent, (also called biodegradable, or naturally-dissolving) serves the same purpose, but is manufactured from a material that may dissolve or be absorbed in the body.

Background

The use of metal drug-eluting stents presents some potential drawbacks. These include a predisposition to late stent thrombosis, prevention of late vessel adaptive or expansive remodeling, hindrance of surgical revascularization, and impairment of imaging with multislice CT.[1][2]

To overcome some of these potential drawbacks, several companies are pursuing the development of bioresorbable or bioabsorbable stents. Like metal stents, placement of a bioresorbable stent will restore blood flow and support the vessel through the healing process. However, in the case of a bioresorbable stent, the stent will gradually resorb and be benignly cleared from the body, leaving no permanent implant.

Studies have shown that the most critical period of vessel healing is largely complete by approximately three months.[3][4] Therefore, the goal of a bioresorbable or “temporary” stent is to fully support the vessel during this critical period, and then resorb from the body when it is no longer needed.

Base materials

Bioabsorbable scaffolds, or naturally dissolving stents, that have been investigated include base materials that are either metals or polymers. Those that have been approved in markets around the world and thus have gained the most traction are based on polymers that are similar to those used in dissolvable stitches.

Metal based

Metal stent candidates are iron,magnesium, zinc and their alloys.

Iron stents were shown using an in vivo evaluation method based on the murine abdominal aorta to generate an iron oxide-filled cavity in the vascular wall.[5] This behavior significantly narrowed the lumen and generated a potential site for rupture of the endothelium after stent degradation.

Magnesium is a relatively new biomaterial that has recently been gaining traction.[6] While degrading harmlessly, it has been shown to possess a functional degradation time of about 30 days in vivo. This is much short of the three-to-six month window desired for bioabsorbable stents. Thus, much attention has been given to drastically reducing the rate of magnesium corrosion by alloying, coating, etc.[7] Many novel methods have surfaced to minimize the penetration rate and hydrogen evolution rate (or, in layman's terms, the corrosion rate). One of the most successful has involved the creation of bioabsorbable metallic glasses via rapid solidification. Other, alternative solutions have included the development of magnesium-rare earth (Mg-RE) alloys which benefit from the low cytotoxicity of RE elements. Coatings and sophisticated materials processing routes are currently being developed to further decrease the corrosion rate. However a number of issues remain limiting the further development of Mg biomaterials in general.[8]

Recently, zinc was shown to exhibit outstanding physiological corrosion behavior, meeting a benchmark penetration rate of 20 micrometers per year.[9] This contribution also asserts that zinc alloys generally meet or exceed mechanical behavior benchmarks (i.e. ductility and tensile strength). While promising, this material is relatively new, so further work is required to prove that zinc is a feasible base material for a stent.

Polymer-based

Polymer-based stents have been approved for use in some countries around the world. These are based on poly(L-lactide) (PLLA), chosen because it is able to maintain a radially strong scaffold that breaks down over time into lactic acid, a naturally occurring molecule that the body can use for metabolism. Other polymers in development include tyrosine poly carbonate and salicylic acid.[10]

An example of a naturally dissolving stent is the 'Absorb' stent 'produced by Abbott that has several design components and features: base scaffold: a poly(L-lactide) polymer similar to that in dissolvable stitches is shaped into a tube made up of zigzag hoops linked together by bridges; drug-eluting layer': a mixture of poly-D, L-lactide (PDLLA) and everolimus; 'markers': a pair of radio-opaque platinum markers at the ends that allow the device to be visualized during angiography; 'delivery system': a balloon delivery system.

Clinical Research

Clinical research has shown that resorbable scaffolds, or naturally dissolving stents, offer comparable efficacy and safety profile to drug-eluting stents. Specifically, the Absorb naturally dissolving stent has been investigated in single-arm trials and in randomized trials comparing it to a drug-eluting stent (DES). Early and late major adverse cardiac events, revascularizations, and scaffold thromboses have been uncommon and similar to the Xience DES, a market leader in the drug eluting stent category.[11][12][13][14][15] Studies in real-world patients are ongoing.[15]

Imaging studies show that the Absorb naturally dissolving stent begins to dissolve from six to 12 months and is fully dissolved between two to three years after it is placed in the artery.[13] Two small platinum markers remain to mark the location of the original PCI. The artery is able to dilate and contract, called vasomotion, similar to a healthy blood vessel at two years.[12]

See also

References

  1. Serruys, PW; Ormiston JA; Onuma Y; et al. (14 March 2009). "A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods". Lancet. 373 (9667): 897–910. doi:10.1016/S0140-6736(09)60325-1. PMID 19286089.
  2. Ormiston, JA; Serruys PW; Regar E; et al. (15 March 2008). "A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial". Lancet. 371 (9616): 899–907. doi:10.1016/S0140-6736(08)60415-8. PMID 18342684.
  3. Serruys, PW; Luijten HE; Beatt KJ; et al. (February 1988). "Incidence of restenosis after successful coronary angioplasty: a time-related phenomenon. A quantitative angiographic study in 342 consecutive patients at 1, 2, 3, and 4 months.". Circulation. 77 (2): 361–71. doi:10.1161/01.CIR.77.2.361. PMID 2962786.
  4. Post, MJ; Borst C; Kuntz RE (1994). "The relative importance of arterial remodeling compared with intimal hyperplasia in lumen renarrowing after balloon angioplasty: a study in the normal rabbit and the hypercholesterolemic Yucatan micropig". Circulation. 89 (6): 2816–2821. doi:10.1161/01.CIR.89.6.2816. PMID 8205696.
  5. Pierson, D; Edick J; Tauscher A; Pokorney E; Bowen PK; Gelbaugh JA; Stinson J; Getty H; Lee CH; Drelich J; Goldman J (January 2012). "A simplified in vivo approach for evaluating the bioabsorbable behavior of candidate stent materials". J Biomed Mater Res Part B. 100B (1): 58–67. doi:10.1002/jbm.b.31922. PMID 21905215. Retrieved 12 October 2012.
  6. Kirkland, N; Birbilis N (2013). Magnesium Biomaterials: Design, Testing and Best Practice. New York: Springer. ISBN 978-3-319-02123-2. Retrieved 2013. Check date values in: |access-date= (help)
  7. Li, N; Zheng Y (2013). "Novel magnesium alloys developed for biomedical application: a review". Journal of Materials Science & Technology. ISBN 978-3-319-02123-2.
  8. Kirkland, Nicholas T.` (2012). "Magnesium biomaterials: past, present and future". Corrosion Engineering, Science and Technology. doi:10.1179/1743278212Y.0000000034.
  9. Bowen, PK; Drelich J; Goldman J (14 March 2013). "Zinc Exhibits Ideal Physiological Corrosion Behavior for Bioabsorbable Stents". Advanced Materials. 25 (18): 2577–82. doi:10.1002/adma.201300226. PMID 23495090. Retrieved 15 March 2013.
  10. Gogas BD, Farooq V, Onuma Y, Serruys PW (2012). "The ABSORB bioresorbable vascular scaffold: an evolution or revolution in interventional cardiology?" (PDF). Hellenic J Cardiol. 53 (4): 301–309. 22796817.
  11. Ormiston JA, Serruys PW, Regar E; et al. (2008). "A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial". Lancet. 371 (9616): 899–907. doi:10.1016/S0140-6736(08)60415-8. 18342684.
  12. 1 2 Serruys PW, Ormiston JA, Onuma Y, et al. A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging methods. Lancet 2009;373(9667):897-910
  13. 1 2 Serruys PW, Onuma Y, Garcia-Garcia HM, et al. Dynamics of vessel wall changes following the implantation of the absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study at 6, 12, 24 and 36 months. EuroIntervention 2014;9(11):1271-1284
  14. Serruys PW, Chevalier B, Dudek D; et al. (2015). "A bioresorbable everolimus-eluting scaffold versus a metallic everolimus-eluting stent for ischaemic heart disease caused by de-novo native coronary artery lesions (ABSORB II): an interim 1-year analysis of clinical and procedural secondary outcomes from a randomised controlled trial". Lancet. 385 (9962): 43–54. doi:10.1016/S0140-6736(14)61455-0. 25230593.
  15. 1 2 Smits P, Ziekenhuis M, Absorb Extend: an interim report on the 36-month clinical outcomes from the first 250 patients enrolled. Presented at Transcatheter Cardiovascular Therapeutics (TCT) conference 2014 in Washington, DC, September 2014
This article is issued from Wikipedia - version of the 11/27/2016. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.